Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use – Part 2-081: Particular Requirements for Automatic and Semi-Automatic Laboratory Equipment for Analysis and Other Purposes
UL 61010-2-081 (IEC 61010-2-081:2015)
1 Scope and object
This clause of Part 1 is applicable except as follows:
1.1.1 Equipment included in scope
Replacement: – Replace the text by the following: This part of IEC 61010 applies to automatic and semi-automatic equipment for analysis and other purposes.Automatic and semi-automatic laboratory equipment consists of instruments or systems for measuring or modifying one or more characteristics or parameters of samples, performing the complete process or parts of the process without manual intervention. Equipment forming part of such a system is within the scope of this standard.Examples of equipment within the scope of this standard include:
NOTE 1 In the case of analytical equipment the complete process usually includes the following steps:
NOTE 2 If all or part of the equipment falls within the scope of one or more other part 2 standards of IEC 61010 as well as within the scope of this standard, considerations have to be given to those other part 2 standards.
1.1.2 Equipment excluded from scope
Addition: Add the following item:aa) IEC 61010-2-101 (IVD Equipment)
1.2 Object
1.2.1 Aspects included in scope
Addition: Add the following items:aa) biohazards;bb) hazardous chemical substances.
1.2.2 Aspects excluded from scope
Addition: Add the following item and note: aa) handling or manipulation of material outside the equipment.
NOTE Requirements covering these subjects are the responsibility of committees preparing relevant standards.