1.1 This test method covers the determination of low amounts of potassium in waters2 having low solids content. The applicable range of this test method is 0.20 to 4.0 mg/L when using the 766.5-nm resonance line. The range may be extended upward by dilution of an appropriate aliquot of sample or by using the less-sensitive 404.4-nm resonance line. Many workers have found that this test method is reliable for potassium levels to 0.02 mg/L, but use of this test method at this low level is dependent on the configuration of the aspirator and nebulizer systems available in the atomic absorption spectrophotometer as well as the skill of the analyst. The precision and bias data presented are insufficient to justify use of this test method in the 0.02-mg/L range.
Product Details
- Published:
- 02/01/2015
- Number of Pages:
- 4
- File Size:
- 1 file , 78 KB
- Redline File Size:
- 2 files , 150 KB